mindlabs
Solutions
8 Solutions

IoT Monitoring Solutions

Environmental monitoring across critical industries.

Cold Chain & Warehouse Monitoring

Multi-zone tracking for cold storage, warehouses & reefer vehicles

Cold Storage Monitoring System

Real-time monitoring for cold rooms, freezers & warehouses

Environmental Monitoring System in Pharma & Clean Rooms

Cleanroom & controlled environments

Temperature Monitoring System for ILRs & Freezers

ILRs, freezers, pharmacy & vaccine storage

Hospital Environmental Monitoring System

Pharmacies, vaccine fridges, blood banks, labs, OTs & ICUs

Pharmaceutical Temperature Monitoring System

Pharma warehouses, labs and shipments

IoT Data Loggers for OCEMS

Industrial emission monitoring

Server Room & Data Center Temperature & Humidity Monitoring

Rack-level sensors, ASHRAE-ready dashboards

Products
4 Products

Temperature Data Loggers

IoT data loggers for real-time environmental and shipment monitoring.

View all products
Anchor

Anchor

Real-time temp & humidity monitoring with 4G/Wi-Fi & 1-year battery.

Go

Go

In-transit temp & GPS tracking for exports, air cargo & domestic delivery.

Lite Plus

Lite Plus

Plug-and-play USB logger for offline temperature recording & PDF reports.

Nova

Nova

Offline multi-use logger for validation studies, temperature mapping & long-term monitoring.

Resources
Resources

Learn & Download

Guides, blogs and datasheets.

View all guides

Guides

Step-by-step buying and compliance guides

Blogs

Latest insights and articles

Case Studies

How customers use Mindlabs monitoring

Datasheets

Product specs to download

Company
Company

About Mindlabs

Who we are and how to join us.

See open roles

About

Our story, team and mission

Careers

Open roles in Hyderabad

Contact

Ready to Transform your Monitoring?

IoT-powered monitoring that ensures quality, compliance, and control across storage, manufacturing, and transit.

Level 2, Trendz Inspire, Kavuri Hills Rd,

CBI Colony, Madhapur, Hyderabad,

Telangana 500033

sales@mindlabs.cloud+91 92475 58750+91 92479 14288

Quick Links

  • Solutions
  • Products
  • About
  • Careers
  • Contact
  • FAQs
  • Blogs
  • Guides
  • Datasheets

Products

  • Anchor
  • Go
  • Lite Plus
  • Nova

Tools

  • Sensor Drift Risk Score

Solutions

  • Cold Chain & Warehouse Temperature Monitoring System
  • Cold Storage Monitoring
  • Environmental Monitoring System for Pharma Cleanrooms, GMP & Labs
  • Temperature Monitoring System for ILRs & Freezers
  • Hospital Environmental Monitoring System
  • Pharmaceutical Temperature Monitoring System
  • FMCG & Food Temperature Monitoring System
  • IoT Data Loggers for OCEMS
  • Server Room & Data Center Temperature & Humidity Monitoring System
  • Temperature Data Logger Manufacturers in India
  • 21 CFR Part 11 Compliant Temperature Monitoring System

Copyright © 2026 Alomind Labs Private Limited

|
All Rights Reserved
|Terms and Conditions|Privacy Policy|Code of Conduct
Chat on WhatsApp
HomeGuidesIQ OQ PQ in pharma

Compliance guide

IQ OQ PQ in Pharma: A Practical Guide to Installation, Operational and Performance Qualification

Every stability chamber, cold room, freezer and controlled-ambient store in a pharma plant has to pass three documented stages before it holds product: installation qualification (IQ), operational qualification (OQ) and performance qualification (PQ). Revised Schedule M made this explicit for Indian manufacturers in December 2023, in line with WHO GMP and EU GMP Annex 15.

This guide shows what each stage tests, the numbers inspectors expect to see, and the documents your qualification file needs. It focuses on temperature- and humidity-controlled equipment and areas, where a failed qualification puts stored product at risk.

Mindlabs Team Updated 11 Nov 2026 20 min read

At a glance

Written for
QA, validation, engineering and QC teams
Covers
12 chapters
Reading time
20 minutes
Start reading
IQ OQ PQ qualification of a stability chamber in a pharma plant

Quick answer

  • IQ proves it was installed as specified. OQ proves it works as specified when empty. PQ proves it holds conditions when loaded and in daily use.
  • Each stage must be completed, and its deviations closed, before the next one starts.
  • WHO expects at least 24 hours of empty-chamber data in OQ, and door-opening recovery within a defined time (for example 30 minutes for a 2–8 °C store) in PQ.
  • Acceptance criteria come from your URS and are fixed before testing begins.
  • Requalify on a justified schedule and after any significant change.
  • Download: IQ/OQ/PQ protocol templates for chambers, cold rooms and warehouses
On this page
  1. What IQ, OQ and PQ mean
  2. URS, DQ and the full sequence
  3. Difference between IQ, OQ and PQ
  4. Installation Qualification (IQ)
  5. Operational Qualification (OQ)
  6. Performance Qualification (PQ)
  7. Acceptance criteria
  8. Protocol and report
  9. Requalification
  10. Monitoring system and loggers
  11. Eight qualification mistakes
  12. FAQ

What IQ, OQ and PQ mean in pharma

IQ, OQ and PQ are the three documented stages that prove equipment or a facility is fit for its intended GMP use. The definitions below follow WHO's technical supplement on qualifying temperature-controlled storage.

StageFull formQuestion it answersCondition during testing
IQInstallation QualificationWas it built and installed to the design specification?Installed, not yet operated for GMP use
OQOperational QualificationDoes it operate to specification under controlled conditions?Empty
PQPerformance QualificationDoes it perform consistently in routine use?Loaded, normal operations
5.2°IQOQPQInstalledRuns emptyLoaded, in use

Qualification vs validation. Qualification applies to premises, utilities and equipment. Validation applies to processes, methods and computerised systems. Qualified equipment is a precondition for a validated process; qualification alone does not validate the process.

Where the requirement comes from

Regulation or guidelineWhat it says on qualification
Revised Schedule M (G.S.R. 922(E), 28 Dec 2023)Section 5 extends qualification to premises, utilities, equipment and process (DQ, IQ, OQ, PQ). Section 20 requires IQ and OQ of computerised system hardware and software. In force for large manufacturers from 28 June 2024. Schedule M guide.
EU GMP Annex 15 (in force 1 Oct 2015)Lifecycle from URS through DQ, FAT/SAT, IQ, OQ, PQ. IQ and OQ may be combined. Adds a requalification chapter. A revision concept paper was published in 2026.
WHO TRS 961 Annex 9, Supplement 7 (TRS 992 Annex 5, 2015)Every new temperature-controlled store must be qualified before release for routine storage of temperature-sensitive products.
US FDA 21 CFR 211Equipment must be of appropriate design and suitably located; qualification is the evidence inspectors review.
The India angle: until the 2023 revision, Schedule M covered process, testing and cleaning validation but had no explicit equipment qualification provision. Plants preparing for revised Schedule M inspections now need a complete DQ-to-PQ file for each critical item.

Where URS and DQ fit: the full qualification sequence

URS template (PDF) Traceability matrix (PDF)

IQ, OQ and PQ test against requirements written earlier. If those requirements are missing or vague, there is nothing to test against. Annex 15 sets the order below; WHO says each stage must be completed before the next begins.

  1. 1URS (User Requirements Specification): what the equipment must do, written by the user. For a stability chamber: setpoint 30 °C / 65% RH, tolerance ±2 °C / ±5% RH, alarms, data retention, Part 11 needs.
  2. 2DQ (Design Qualification): documented check that the proposed design meets the URS and GMP. Done before purchase order or construction.
  3. 3FAT (Factory Acceptance Test): optional; tests at the supplier before shipping. Annex 15 recommends it for novel or complex equipment.
  4. 4SAT (Site Acceptance Test): optional; checks after delivery. Justified FAT/SAT results can be reused in IQ/OQ if transport and installation did not affect them.
  5. 5IQ, then OQ, then PQ.
  6. 6Release for routine use, then periodic review and requalification.
Verified byURSFunctional specDesign specDQIQOQPQ

The V-model pairing (each test stage verifies one specification level):

SpecificationVerified byExample for a 2–8 °C cold room
URSPQHolds 2–8 °C everywhere product is stored, loaded, in daily use
Functional specificationOQController holds setpoint empty; alarms fire at limits; recovers after power loss
Design specificationIQRefrigeration unit model, panel build, sensor count and wiring match drawings

The link between them is the traceability matrix: every URS line maps to at least one test with a pre-set acceptance criterion. Auditors check it to confirm nothing in the URS went untested.

Difference between IQ, OQ and PQ

The three stages differ in what they test against, the condition of the equipment, and who usually signs. A 2–8 °C cold room is used as the example throughout.

IQOQPQ
Tests againstDesign specification and drawingsFunctional specificationURS (intended use)
Equipment stateInstalled, not operating for GMP useRunning, emptyRunning, loaded as in normal use
Typical checksModel and serial numbers, utilities, wiring, calibration certificates, manuals, spare partsController and setpoints, high/low alarms, empty mapping, power failure, SOPs and trainingLoaded mapping, door-opening recovery, seasonal performance
Typical duration1–3 days of inspectionAt least 24 h of empty data, plus alarm and power testsSeveral days of loaded data; often repeated in summer and winter
Cold room pass exampleRefrigeration unit matches specified model and capacityAll loggers stay within 2–8 °C for 24 h emptyAll loggers return to 2–8 °C within 30 min of door closing
Typical executorsEngineering, vendor, validationValidation, engineeringValidation, QA, operations
OutputIQ report; deviations closed before OQOQ report; deviations closed before PQPQ report; equipment released for routine use

IQ and OQ can be combined into a single IOQ protocol where justified (Annex 15). WHO notes that OQ and PQ for storage areas are often run as one sequence after IQ.

Installation Qualification (IQ): what to verify

IQ protocol (PDF) Deviation report (PDF)

IQ confirms that what was installed matches what was specified, item by item. Every check is recorded in three columns: specified, as found, and pass or fail. Fill in the "specified" column before the inspection starts, so the inspector is checking against the design and not against what happens to be there.

Critical components to list first

  • Refrigerators, freezers, small chambers: thermostat or controller, its control sensor, and the independent monitoring device and its sensor.
  • Cold rooms, freezer rooms, walk-in chambers: controller, sensors, cooling unit, condenser, evaporator, and the room enclosure itself (panels, doors, seals). Where components are duplicated, check each one.
4.6°EvaporatorControl sensorMonitoring loggerControllerDoor sealsCondensing unitInsulated panels

IQ checklist

AreaWhat to checkExample of a fail
IdentityManufacturer, model, serial number, internal asset IDUnit installed is a lower-capacity model than the one specified
Electrical supplyVoltage (e.g. 415 V, 3-phase, 50 Hz), circuit amperage, breaker label, earthing, isolating switch, emergency powerCircuit rated 20 A where 30 A was specified
EnvironmentCleanliness, fumes, vibration, ambient temperature and RH at handoverRefrigeration unit vibrating on its mount
CalibrationCurrent certificate for every critical controller and sensor, traceable to national standards; tag on eachSensor certificate expired or not traceable
Software and firmwareVersions recorded, configuration backed upFirmware version differs from the one in the design documents
DocumentsDrawings, manuals, wiring diagrams, material certificatesAs-built drawings missing
Spare partsList supplied parts and check them against specificationSpare sensor is a different type from the installed one
Auxiliary equipmentGenerator, UPS, building alarm, network; record the interfacesGenerator changeover not documented
Maintenance readinessAsset in equipment inventory, labelled, maintenance logbook in placeNo preventive maintenance entry created
IQ data sheetSpecifiedAs foundDEV-014Specified · As found · Pass / Fail

Two practices auditors look for

  • A calibration check straight after installation, before OQ starts, so transport or installation has not shifted any sensor.
  • Every deviation on a separate deviation form, cross-referenced from the IQ sheet, with a corrective action and an owner. WHO's rule: do not begin OQ until relevant IQ deviations are corrected.

All sheets are signed by the inspector and checked by a reviewer who did not perform the inspection.

Operational Qualification (OQ): the tests and what passes

OQ protocol (PDF)

OQ proves the equipment works as specified while it is empty. WHO's rule: do not start OQ until the IQ deviations are corrected. Run the tests below in this order, using the same logger layout throughout so results can be compared.

#OQ testHow it is doneTypical acceptance criterion
1Calibration statusCheck every critical controller and sensorCurrent certificate, traceable to national standards, tagged with ID, cal date and due date, entered in the calibration programme
2SOPs and trainingCheck operation, alarm response and maintenance SOPs existSOPs approved and matched to the installed equipment; training record for each SOP
3Control panel and setpointsExercise controls, displays and setpoint changesAll functions work per manufacturer specification
4Alarm testsTrigger at least one high and one low temperature alarm, plus door-open and power alarms if fittedAlarm activates on every configured channel (sounder, SMS, email, call); settings recorded
5Empty-chamber distribution studyLoggers at defined points, empty, at setpointAll points within limits for the whole run; WHO minimum is 24 hours
6Door-opening recovery (empty)Open the door for the longest routine durationReturns within limits inside the time set in the URS
7Power failureCut power until out of range, or for a fixed period (e.g. 2 hours), then restoreHoldover time and recovery time recorded; pass/fail only if the URS sets a limit
1Calibration2SOPs3Controls4Alarms524 h empty6Door7Power

Notes inspectors raise

  • Alarm tests must reach a person. An alarm that sounds on the panel but never sends the SMS fails, because no one acts on it at 2 a.m. Test every escalation contact listed in the alarm SOP.
  • Power failure results are design data. WHO sets no fixed pass mark for large stores. The holdover time tells you how long you have before product is at risk and how big your generator or UPS must be. Write it into the excursion SOP.
  • One empty study is enough for OQ. WHO recommends seasonal studies for full characterisation, but only one empty run is required at OQ stage.
  • Logger specification matters. WHO lists the minimum for qualification loggers: a range covering the extremes (for example −30 °C to +60 °C), a programmable interval of 1–15 minutes, a 3-point traceable calibration certificate with error no worse than ±0.5 °C, and Part 11-capable software.
2°C8°COffOnHoldoverRecovery9.6°

How many loggers, where to place them and how to analyse the data is covered in our temperature mapping guide. This chapter only sets out what OQ must prove.

Performance Qualification (PQ): proving it in routine use

PQ protocol (PDF)

PQ proves the equipment holds its conditions loaded, with doors opening and people working around it. WHO's rule: start PQ only after OQ deviations are corrected, and release the equipment only after PQ deviations are corrected.

What PQ covers

PQ testHow it is doneTypical acceptance criterion
Calibration re-checkConfirm controllers and sensors are still in calibrationAll current; no drift beyond tolerance since OQ
Loaded distribution studySame logger layout as OQ, with product or a representative load at normal or maximum loadingEvery point within the specified range for the full run
Door-opening recovery (loaded)Open the door(s) for the longest duration seen in routine work; with several doors, use the real sequenceAll loggers back within range in the URS time; WHO's example is 2–8 °C within 30 minutes of closing
Each operating conditionFor multi-setpoint stability chambers, repeat at every condition in use (e.g. 25 °C/60% RH, 30 °C/65% RH, 30 °C/75% RH, 40 °C/75% RH)Within ±2 °C and ±5% RH of each setpoint
Seasonal performance (areas affected by ambient)Repeat the loaded study in the hottest and coldest periodsWithin range in both. In most of India that means April–June and December–January; check RH in the monsoon too
2°C8°COpenClosed≤ 30 min

Three PQ decisions to fix in the protocol, not during the run

  1. 1Load pattern. State the load as a percentage of capacity and how it is arranged. Inspectors ask whether PQ was run at worst-case (maximum) load.
  2. 2Door-opening duration. Base it on observed behaviour, such as your longest receiving or pick event, and record where that number came from.
  3. 3Study duration. State the hours or days and the logging interval before starting. A study cannot be cut short because the data look good.

After PQ, the qualified state has to be maintained. Monitoring sensors drift between qualifications; see our blog on why monitoring drifts after validation.

Writing qualification acceptance criteria (with typical values)

Acceptance criteria must be numeric, written into the protocol and approved before testing starts. Changing them after you see the data invalidates the qualification and is a common inspection finding. Derive each limit from three inputs: the product's labelled storage condition, the URS, and the equipment's specified capability.

Typical starting values by equipment type (confirm against your own URS and product labels)

EquipmentTypical rangeStability or uniformity criterionRecovery criterion
Stability chamber, long-term (India, climatic zone IVb)30 °C / 75% RHEvery point within ±2 °C and ±5% RH (ICH tolerance)Back within tolerance in the URS time after a door opening
Stability chamber, accelerated40 °C / 75% RHEvery point within ±2 °C and ±5% RHAs above
Stability chamber, intermediate or zone II long-term30 °C / 65% RH or 25 °C / 60% RHEvery point within ±2 °C and ±5% RHAs above
Pharmaceutical refrigerator, cold room2–8 °C (ICH refrigerated long-term: 5 °C ± 3 °C)Every point within 2–8 °C; no reading at or below 0 °CWHO example: within range 30 minutes after door closes
Freezer−25 °C to −10 °C (USP freezer range)Every point within rangeDefined in URS
Ultra-low freezer−80 °C or −70 °C setpoint, per productEvery point at or below the product limitDefined in URS; record holdover time
Controlled-ambient warehousePer label, e.g. 15–25 °C or "store below 30 °C"Every point within the label range, in hot and cold seasonsDock-door recovery defined in URS
30°·75%30.4°Temperature30 °C±2 °CHumidity75% RH±5%

Rules for writing a criterion that holds up

  • Write a number, not a word. "Within 2.0–8.0 °C at all 15 points for 24 hours" is testable. "Temperature uniform" is not.
  • Use the same precision as your loggers. With ±0.5 °C loggers, a criterion of ±0.2 °C cannot be proved.
  • Say what happens at the edge. State whether 8.0 °C is a pass, and whether a single reading out of range fails the study or triggers a deviation.
  • Don't copy criteria between rooms. A protocol reused from another facility without re-deriving limits from its own URS is a common and serious failure.

IQ OQ PQ protocol and report: the documents auditors check

Qualification plan (PDF) Summary report (PDF)

An inspector reviewing a qualification asks for one file per equipment item. If any document below is missing, unsigned or out of sequence, expect a finding.

Every IQ, OQ and PQ protocol needs seven sections (WHO structure):

  1. 1Approval page and change history (author, reviewer, approver, version log)
  2. 2Acronyms and glossary
  3. 3Description and rationale: what is being qualified and why
  4. 4Scope and objectives
  5. 5Key parameters: setpoints, ranges, alarm limits
  6. 6Test procedures, each with title, target, step-by-step method, acceptance criterion and a blank data sheet
  7. 7Report template

The qualification file

DocumentWhen it is createdWhat auditors check
URSBefore purchaseSpecific, testable requirements; approved by QA
DQ reportBefore purchase or buildEach URS line traced to the design
Validation master plan or qualification planBefore IQScope, sequence, responsibilities
IQ, OQ, PQ protocolsApproved before each stageApproval date earlier than execution date; criteria pre-set
Executed data sheetsDuring testingSpecified, as found, pass/fail; signatures and dates; raw logger files attached
Deviation reportsWhenever a result failsImpact assessment, corrective action, closure before the next stage
Calibration certificatesBefore OQEvery logger and every control or monitoring sensor, traceable, in date
Traceability matrixUpdated through the projectEvery URS line covered by a test
IQ, OQ, PQ reportsAfter each stageResults summary, deviations, conclusion; reviewed by someone who did not execute
Release for routine useAfter PQQA sign-off with date
URSDQIQOQPQPDFDeviationsCalibration certsTraceability matrixQA release

Requalification: triggers and how often

Requalification tracker (PDF)

No regulation sets a single requalification interval. Annex 15 requires equipment to be evaluated at an appropriate frequency to confirm it stays in a state of control. Where a fixed period is used, it must be justified with defined criteria. WHO adds that qualification is repeated whenever the installation or the way it is used changes significantly.

Event-based triggers (requalify the affected stage)

ChangeUsual scope
Relocation of the equipmentFull IQ, OQ and PQ
Replacement of controller, compressor, evaporator or control sensorIQ of the new part, then OQ; PQ if performance could be affected
Firmware or software change in the controller or monitoring systemChange control impact assessment; OQ of affected functions
New setpoint or operating rangeOQ and PQ at the new condition
Change in load pattern, racking or layoutPQ
Recurring excursions or an alarm trendInvestigation, then targeted OQ or PQ
Major repair, panel or door replacement in a cold roomIQ of the repaired element, then OQ
IQOQPQChange or eventImpact assessmentRepeat IQ / OQ / PQPeriodic reviewQualified state

Time-based review (common industry practice, not a legal number)

  • Stability chambers: OQ and PQ are usually repeated every year.
  • Cold rooms and warehouses: often every 1–3 years, set by risk and justified in the qualification plan.
  • Continuously monitored, stable equipment: a documented periodic review of alarm history, deviations, calibration and maintenance can justify a longer interval than a fixed calendar repeat.

For a worked example of setting the interval in a regulated lab, see How Often Should You Requalify a CRO Freezer?

Qualifying the monitoring system and the loggers you test with

Two things are often missed: the continuous monitoring system on the equipment needs its own qualification, and the loggers used in OQ and PQ need to be fit for the job.

The monitoring system is a computerised system. Revised Schedule M (Section 20) requires IQ and OQ of computerised system hardware and software. For a temperature and humidity monitoring system that typically means:

StageWhat to verify
IQSensor and gateway models, firmware versions, installed locations matching the drawing, calibration certificate for each sensor, network and power connections, user accounts created
OQHigh and low alarms reach every contact; time sync; audit trail records setpoint and limit changes; role-based access; data backup and restore; report output matches raw data
PQRuns in routine use for a defined period with real users responding to real alerts
30.1°4.8°-21°IQOQPQSMS · email · callSensorsDashboard + audit trail

The electronic records rules (audit trails, e-signatures, access control) are covered on our 21 CFR Part 11 temperature monitoring page.

Loggers used for qualification studies. Check them against the WHO minimum before the study, not after:

  • Range covering the expected extremes, for example −30 °C to +60 °C
  • Programmable logging interval of 1–15 minutes, with memory for the full study
  • 3-point calibration certificate traceable to national standards, error no worse than ±0.5 °C
  • Data download to software that meets 21 CFR Part 11

Mindlabs Nova is built for qualification and validation studies: ±0.5 °C accuracy, temperature and humidity, a configurable logging interval, 20 months of recording and auto-generated PDF/CSV reports. It ships with a calibration certificate; with the PT-100 probe it logs down to −80 °C for ultra-low freezers. For continuous monitoring after release, Anchor sends real-time alerts.

Mindlabs Nova data logger used for qualification studies

Request a free Nova trial for your next qualification.

Request a free trial

Eight qualification mistakes that become inspection findings

Most qualification findings are not about the equipment. They are about missing proof. The FDA has cited a stability chamber qualification in a warning letter because the studies did not always include specifications, acceptance criteria or raw data. In FDA observation counts presented in 2019, "calibration/inspection/checking not done" (21 CFR 211.68(a)) was cited 60 times and "equipment design, size and location" (211.63) 53 times.

#MistakeWhat the inspector seesFix
1Protocol approved after testing beganApproval date later than first data sheetApprove every protocol before execution
2Acceptance criteria missing, vague or changed later"Uniform" instead of a number; revised limits after resultsNumeric criteria in the approved protocol; changes only through change control before the test
3Generic protocol copied from another siteLimits that do not match this room's URSRe-derive limits from this equipment's URS
4OQ started with IQ deviations openDeviation closure dated after OQ startClose or formally risk-assess deviations before the next stage
5Expired or untraceable calibrationLogger certificate out of date on the study dateCheck every certificate before the study starts and attach it to the report
6Alarms tested only at the panelNo proof SMS or email reached anyoneTest every notification channel and contact
7PQ run only empty, or at partial loadNo loaded or worst-case dataDefine the load pattern and run PQ loaded
8No requalification after a changeCompressor replaced, no change control or retestLink maintenance work orders to change control and requalification
Data sheet12 MarProtocol approval18 MarApproved after testing
A ninth gap common in Indian plants preparing for revised Schedule M: equipment installed years ago with no IQ on file. Retrospective qualification is discouraged; the practical route is a documented gap assessment, then current-state IQ, OQ and PQ.

FAQ

Installation Qualification, Operational Qualification and Performance Qualification. Together with Design Qualification (DQ), they are the four stages of equipment qualification.
Qualification proves premises, utilities and equipment are fit for use. Validation proves a process, method or computerised system consistently gives the expected result. Qualified equipment is a prerequisite for process validation.
Yes. Revised Schedule M (December 2023) extends qualification to premises, utilities and equipment through DQ, IQ, OQ and PQ, and requires IQ and OQ of computerised system hardware and software.
Yes, where justified. EU GMP Annex 15 allows combined IOQ documents. PQ can also be combined with OQ in some cases, but the tests and acceptance criteria for each stage must still be present.
IQ is usually a few days of inspection. OQ needs at least 24 hours of empty-room data plus alarm, door and power tests. PQ needs a loaded study, and seasonal studies if the room is affected by ambient conditions.
Yes, if your QA reviews and approves them and they cover your URS. Annex 15 allows supplier documentation and FAT/SAT results where justified. WHO requires a protocol prepared by a third party to be authorised by the responsible person in your organisation.
The executor, an independent reviewer who did not perform the tests, and QA for final approval. WHO requires results to be reviewed by someone not involved in the exercise.
Yes, if they store GMP materials or temperature-sensitive products. WHO allows a reduced sequence for small cabinets but still expects measurements at top, bottom, front and back. Domestic refrigerators are not recommended for vaccines.
After any significant change, and periodically at a justified interval. Stability chambers are commonly requalified every year; storage areas every 1–3 years based on risk.

On this page

0% read
  1. What IQ, OQ and PQ mean
  2. URS, DQ and the full sequence
  3. Difference between IQ, OQ and PQ
  4. Installation Qualification (IQ)
  5. Operational Qualification (OQ)
  6. Performance Qualification (PQ)
  7. Acceptance criteria
  8. Protocol and report
  9. Requalification
  10. Monitoring system and loggers
  11. Eight qualification mistakes
  12. FAQ

Need help with this?

Our team can review your requirements and suggest the right monitoring setup.

Talk to the team
DatasheetsAll guides

More guides

All guides
Buying GuidesHow to Choose a Temperature Data Logger in IndiaMatch a logger to your use case, alerts, range and accuracy, compliance records and total cost, with a URS checklist you can hand to vendors. 15 min readComplianceRevised Schedule M Compliance GuideWhat revised Schedule M changed, the deadlines, what inspectors ask to see, and a compliance checklist with a 90-day compliance plan. 17 min readCold ChainCold Chain SOPs for Pharma: The 10 SOPs You NeedThe 10 SOPs every pharma site handling cold chain products needs, from receipt and storage to excursions, transport, mapping and calibration, with sample templates to adapt. 22 min read